Variable Temperature Sample Environment Summary

Variable Temperature Sample Environment Summary

Equipment

Sample type

Temperature range and operating units

Ramp rate limits and units

Setup time (hrs) (includes normal beamline setup)

Other info

Which column in a .asc file is the temperature?

Serpentine hot air blower

Broken quartz capillary

Room Temperature - 750°C

50°/minute heating or cooling

5

Users should do their own calibration

SR10BM01HU02IOC01:ET:RBV_pvTemp 0 Eurotherm_Temperature_(C)

 

 

 

 

 

 

 

Cyberstar hot air blower

Broken quartz capillary

Room Temperature - 950°C

6°/minute heating, 15°/minute cooling

5

Temperature calibration exists

SR10BM01HU02IOC01:ET:RBV_pvTemp 0 Eurotherm_Temperature_(C)

 

 

 

 

 

 

 

Oxford Cryosystems Cryostream

Broken quartz capillary

80 - 500 K

360°/hour (6°/min)

Time needed to cool from 300 K to 80 K: 1 h.

5

Equipment is calibrated

SR10BM01CS800:CURRENTTEMPERATURE

 

 

 

 

 

 

 

Cryostat

 

Capillary (Aluminium)

10 - 300K

4°/minute heating, 2.5°/minute cooling

Time needed to cool from 300 K to 10 K: 2 h 15 m.

10

Aluminium capillaries required, equipment is calibrated

SR10BM01HU02IOC01:CRY:KELVINA_MONITOR 0 Cryostat_Temperature_(K)

 

 

 

 

 

 

 

Stoe Capillary Furnace

Quartz/Sapphire Capillary (full length - with funnel)

Room Temperature - 1200°C

50°/minute heating or cooling

10

Users should do their own calibration  - Kapton windows add to low angle background 

SR10BM01HU02IOC01:FRNC:RBV_pvTemp 0 Furnace_Temperature_(C)

 

 

 

 

 

 

 

Anton Paar DHS 1100

 

Flat sample/powder

Room Temperature - 1100°C

50°/minute heating or cooling

7

Users should do their own calibration. Corrections to data required due to geometry. Sample chamber can be flooded with inert gas.

SR10BM01HU02IOC01:FRNC:RBV_pvTemp 0 Furnace_Temperature_(C)

 

 

 

 

 

 

 

Anton Paar HTK2000

Flat sample/powder

Room Temperature - 1870 K, using a Pt filament and 2000K with W filament

50°/minute heating or cooling

10

Users should do their own calibration. Corrections to data required due to geometry. Sample chamber can be flooded with inert gas.

SR10BM01HU02IOC01:FRNC:RBV_pvTemp 0 Furnace_Temperature_(C)

 