Variable Temperature Sample Environment Summary

Equipment

Sample type

Temperature range and operating units

Ramp rate limits and units

Setup time (hrs) (includes normal beamline setup)

Other info

Which column in a .asc file is the temperature?

Serpentine hot air blower

Broken quartz capillary

Room Temperature - 750°C

50°/minute heating or cooling

5

Users should do their own calibration

SR10BM01HU02IOC01:ET:RBV_pvTemp 0 Eurotherm_Temperature_(C)

 

 

 

 

 

 

 

Cyberstar hot air blower

Broken quartz capillary

Room Temperature - 950°C

6°/minute heating, 15°/minute cooling

5

Temperature calibration exists

SR10BM01HU02IOC01:ET:RBV_pvTemp 0 Eurotherm_Temperature_(C)

 

 

 

 

 

 

 

Oxford Cryosystems Cryostream

Broken quartz capillary

80 - 500 K

360°/hour (6°/min)

Time needed to cool from 300 K to 80 K: 1 h.

5

Equipment is calibrated

SR10BM01CS800:CURRENTTEMPERATURE

 

 

 

 

 

 

 

Cryostat

 

Capillary (Aluminium)

10 - 300K

4°/minute heating, 2.5°/minute cooling

Time needed to cool from 300 K to 10 K: 2 h 15 m.

10

Aluminium capillaries required, equipment is calibrated

SR10BM01HU02IOC01:CRY:KELVINA_MONITOR 0 Cryostat_Temperature_(K)

 

 

 

 

 

 

 

Stoe Capillary Furnace

Quartz/Sapphire Capillary (full length - with funnel)

Room Temperature - 1200°C

50°/minute heating or cooling

10

Users should do their own calibration  - Kapton windows add to low angle background 

SR10BM01HU02IOC01:FRNC:RBV_pvTemp 0 Furnace_Temperature_(C)

 

 

 

 

 

 

 

Anton Paar DHS 1100

 

Flat sample/powder

Room Temperature - 1100°C

50°/minute heating or cooling

7

Users should do their own calibration. Corrections to data required due to geometry. Sample chamber can be flooded with inert gas.

SR10BM01HU02IOC01:FRNC:RBV_pvTemp 0 Furnace_Temperature_(C)

 

 

 

 

 

 

 

Anton Paar HTK2000

Flat sample/powder

Room Temperature - 1870 K, using a Pt filament and 2000K with W filament

50°/minute heating or cooling

10

Users should do their own calibration. Corrections to data required due to geometry. Sample chamber can be flooded with inert gas.

SR10BM01HU02IOC01:FRNC:RBV_pvTemp 0 Furnace_Temperature_(C)