Chemical Crystallography - In person collection

Workflow

Relevant User Interfaces

Step 0: Experimental setup

For chemical crystallography collections, there are a few things you should check before starting to collect:

Step 1: Mount a crystal

  1. Close the shutter (Beam Off button on Dataset window ) and then unlock the door. Ensure the detector has moved all the way back from the goniometer. Do not reach around it to mount your sample.

  2. Place your pin carefully on the goniometer. If you bump it the goniometer it will lose its position and will have to be rehomed, which is a slow task and will take up your beamtime.

  3. Do a search and secure to lock up the hutch.

See for more details.

Step 2: Rastering

If you can’t find your crystal on the loop, use to find it by diffraction. Skip this step if you can see your crystal. Rastering is not really necessary on MX1 since the beam is very large.

 

Step 3: Screening

Collect a small wedge of data to see if the crystals are worth collecting, instructions:

Step 4: Strategy

Use the results from screening to calculate the optimum for your collection.

Step 5: Collect dataset

Collect your perfect and beautiful

Step 6: View processing results and solution

Check your results and decide if you want to collect again for higher quality data or to achieve more data.

This can be done via the data processing webpage, but to find more information, you will want to view the initial solution, which is in the CXASAP-Brute folder in the folder for your dataset. See CX - accessing and analysing data on the beamlinefor how to find your data and for information on how to find your CXASAP results.

Later steps

Prepare your structure for publication

  • Import the relevant synchrotron information into your cif, from the autoprocess.cif file generated for each dataset:

  • Reference the synchrotron and relevant software:

 

 

 