Part 1: Laser and Detector Set up
Move the detector out of the way along the rail
Put the DAC into clamp and tighten the thumb screw
Plug spectrometer into laptop via USB
Slide the detector back along the rail so that it is sitting against the opening of the DAC
Turn the key on the laser PSS to Auto
Lock up the hutch
Picture
Part 2: Take fluorescence measurement
Note: This should only need to be done once at the start of the day, unless you want to change the integration time or number of scans to average
Open OceanView on the laptop
Select Spectroscopy Application Wizard
Select Fluorescence
Click Next to select Active Acquisition
Change Scans to Average to 10 and click Next
Tick the box for Add clean peaks to schematic
Turn off the lights in the room or put the box over the laser setup
Click the big lightbulb icon at the top to store the dark spectrum
Click Finish