Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.


Section



Column

/wiki/spaces/SAXS/overviewail in ← Back to Mail Samples & TemplateTemplates



Column



Div
stylefloat: right; position: sticky; top: 100px; background: rgba(253, 253, 239); border: 3px solid #0096d6; padding: 5px; width: 300px;


Section

Page Index

Table of Contents


...

Column


Section

Steel plates

An experiment specific spreadsheet, to be used as sample list, will be sent out to you prior to your scheduled beamtime.

Use the following spreadsheet as sample list for samples mounted on an example for steel plates.  → Link to Google doc!   → Example Document




Expand
titleHow to fill out the google spreadsheet for steel plates

Prepare one separate google document for ALL samples! Prepare single sheets within the document for each plate!

Image Added

  • Enter all the key information in the google spreadsheet. It will help beamline staff avoid duplication, errors and misunderstandings. The spreadsheets aim to be a single, comprehensive location of the critical information we need to run each sample. 
  • Samples in the order listed (from top to bottom). Lines with the sample name field left empty are skipped by the beamline, so that you can have gaps in the list that way.
  • Sample name for each well
  • Information on how many shots per hole. If more than one, provide the spacings in X (horizontal) and/or Y (vertical) in millimetres.
  • indicate sample prioirity in list - we will leave low prioiry samples until last. If you have all your samples in on spreadsheet, put high priority samples at the top and lower priority one towards the bottom. If you have mutiple spreadsheets, keep samples of different priority in different spreadsheets.
  • Gappy or gapless data
  • Exposure time (seconds)
  • Every google doc must have same name as label on plate
  • Note the estimated run time for each plate and design your experiment accordingly, especially in regard to the amount of samplesThis estimate does NOT include sample / plate changeover and mounting times!


Example below or download Example Spreadsheet here: → Example Document

Image Added




For solid samples, please send us prefilled and labeled steel plates. Mount samples on as few plates as possible, as exchange of plates is not currently automated.

Contact the beamline team (contact details can be found here) to arrange pick up of steel plates and drop off of plates with mounted samples. 

Use the following guidelines for labeling and sample preparation: 

Plate dimensions

We have steel plates for sample mounting with the following applications and dimensions:

40-hole plates:

→ for powders and gels - holes filled with sample and covered with tape on both sides

  • 40 square holes,  each 5 x 5 mm square holes, in 8 columns in 5 rows
  • hole pitch (spacing apart) 14 x 14 mm (horizontal x vertical). Designed for mounting/sealing samples with 12 mm wide tape.
  • available in 0.7 and 2.0 mm thickness. If in doubt, 0.7 mm sample thickness is the safer bet
  • two plates can be mounted on the beamline at one time, so currently 80 samples can be run in one automated acquisition.
  • each plate has a laser-engraved number for each well. Label the plate on the same side as the laser engraved sample numbers. 

X-hole plates

→ for larger, solid samples, fibres etc. NOT recommended for powders, gels etc.

  • 10 mm diameter round holes, typically at 20 x 20 mm horizontal x vertical pitch (spacing apart).
  • limited number of plates, use only where needed
  • only one plate at a time can be loaded onto the beamline
  • there are no labels on the plates. Label each plate, and every individual sample on BOTH sides!

Sample preparation

→ when loading samples, use Kapton tape where you need it, and not where you don't:

  • Kapton tape is great for encapsulating samples that need support, such as powders, gels, some fluids or other formats that are not self supporting for analysis and/or transport. In those cases, you'll need to cover the whole sample with tape, often on both sides, to seal and /or hold the sample in place.
  • if there is tape over the part of sample to be analysed, make sure you include at least one corresponding blank sample (i.e. one or two layers of tape as per your sample mount) so that the background scattering from the tape can be measured, and later subtracted from the sample scattering. If you are running more than one plate, please include at least one blank sample per plate.
  • Kapton (and any other) adhesive tape, if in the X-ray beam during analysis, adds considerable background scattering will affect your data quality. If the sample is self supporting and only needs attaching the metal sample plate, keep the tape clear of where you plan to analyse the sample. Just use small peices of tape at the sides or corners of the sample to hold it onto the plate. In this case, please leave one empty hole in each plate, which will be used to measure the instrument background for subtraction from the sample data.

Labeling

  • See above for labeling guidelines for the two main plate format types.
  • Attach a photo for each plate with visible labels, region of interest and description of preferred sample scans. This is particularly useful if there are visual clues staff need to understand when picking out where you want the sample analysed.

Exceptions

On a case by case basis, there might be exceptions to the way we can run specific sample mounts and setups. This can include us mounting bigger or special samples on different plates or sample mounts on site just prior to running the samples or specific sample preparation that needs to be freshly done before the measurement. Please inquire about these opportunities and arrange these ahead of your beamtime in conversation with the beamline team! Contact information can be found here.


...