Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

 Physical Principles

 

In transmission mode we are measuring the attenuation of the X-ray beam as it traverses a sample of thickness x:

I(E) = I0(E)e−μ(E)·x

Image RemovedImage Added

where I0 and I are the intensity of the incident and transmitted X-ray beam, respectively, and μ is the absorption coefficient. In turn, μ depends on the atomic concentration (ρ) and absorption cross-section (σ):

For a sample consisting of different atoms, the total absorption will be given by the sum of the product for each chemical element present in the sample:

The cross section is a physical property of the element, i.e. a constant (you can find the tabulated values at http://csrri.iit.edu/periodic-table.html ). As such, the way you prepare your sample would affect the absorption coefficient through the atomic concentration (ρi).

Two aspects have to be considered: the total absorption and the edge jump for the element of interest.

The total absorption of the sample should be such that you have enough transmission through the sample to get a decent signal for the X-ray beam after the sample (i.e. in the second ion chamber, I).

A correct edge jump of approximately 1 assures that you can acquire a XAS signal good enough to be analyzed. This is achieved only if the element of interest is a major component because the difference of the absorption coefficient above and under the edge is mainly due to the change in the cross section of the investigated element. In the event the sample is too dilute (by element of interest) you will not be able to acquire a good spectrum because your edge jump will not be big enough even if you obtain the correct value for the total absorption. Adding more product will not help to solve the problem in this case, rather you will need to explore doing the measurement in fluorescence mode.

 

Which kind of sample do I need?

The state of the sample is not particularly important for XAS measurements (it can be a powder, foil, film, amorphous) since this technique probes the local structure around the absorber (no need for a long range ordering). However, certain homogeneity is required.

XAS is a selective technique so you do not have to worry about the presence of other chemical elements in your samples (unless their edges interfere with the investigated edge) but you have to be absolutely confident that your sample does not contain impurities of the investigated chemical element. This is because XAS probes an average local structure around the absorbers.

The most common way to prepare solid samples is to make pellets. For this you will need:

  • The right amount of sample powder fine grained and homogeneous (see next section).

  • A binding/briquetting agent (cellulose is preferred) to add to your sample powder. The agent is added in order to make the pellet more stable and to reach a quantity of total product which is enough make the pellet. For a pellet of approximately 1 cm2 we suggest to use about 100 mg of binding agent.

  • A die set and a hydraulic press (see below: ‘How to make a pellet’).

 

Estimating the right mixture

 In order to calculate properly the quantities necessary for your sample preparation we recommend the use the XAFSmass program :XAFSmass(powders, thin films, foil, glasses, liquids etc.) or the SSRL online calculator (powder samples only):

XAFSMass User guide:

Calculations for different applications are available in this program. However, the most common applications are:

  • the calculation of the quantity (in mg) of powder needed to prepare samples with an optimal absorption for transmission measurements

  • the calculation of thickness and absorption step for samples with known density.

This utility helps you to calculate the quantity (in mg) of powder provided you know:

  1. Sample chemical formula

  2. Desired total absorption above the edge (we suggest to adjust this value to be between 1 and 2 and such that the ‘absorptance step’ calculated by the program would be close to 1)

  3. Sample area

  4. Absorption edge energy

To use XAFSmass to calculate the amount of your target compound required to achieve an edge step of one:

  • Select ‘powder’ from the main menu

  • Type the chemical formula, the desired total absorption (values between 1 and 2 are reasonable), the sample area (this would depend on the die of your hydraulic press; a standard 7mm pellet is 0.38 cm2) and the edge energy of the absorption edge.

  • Press calculate and you can have the needed quantity in mg.

This utility also allows you to calculate the absorptance step for an atom which is substituted in a given percentage in your product. For example if you want the Fe edge in SiO2 with 5 % in weight of Fe you should write in your chemical formula Fe%5SiO_2. Remember that the percentage is given in weight. If you have the atomic percentage than you should use this conversion:

Wt %  = at % * (atomic weight of the chemical element/molecular weight)

 

The ideal absorptance step is 1.
For EXAFS analysis, an edge step >0.3 is generally required.

How to make a pellet

 

  1. Ascertain the amount of quantity of powder you need to prepare your sample (a standard 7mm pellet requires approximately 50 mg cellulose).

  2. Mix your powder sample with a binding agent in a mortar. We suggest to use cellulose (boron nitride diffracts). Ascertain that the two products are homogeneously mixed (use a mortar and pestle or a mini-ball mill).

  3. Use the mechanical auto press or the handheld press (see pictures below). The auto press is located in the chemistry lab, the handheld press is located in the XAS sample mounting area (please use it in the chem. lab. and return after use to the beamline cabin). A manual is available for the auto press in the chem. lab.

Handheld pellet press

Pellet in sample holder

Mechanical autopress

Image RemovedImage Added