Energy Selection for Mail In

When applying for PD Mail in experiments, Users are asked to select an energy at which the measurements will be collected. Available energies are 16 and 21 KeV. These correspond to wavelengths of ~ 0.775 and 0.59Å respectively.

If you still have questions after reading this, please contact the beamline.

Why collect at a particular energy?

There are several reasons to choose a particular energy for measurements and choosing an energy can be a balanced choice between the following factors.

Sample Absorption

Samples with high Z elements will absorb X-rays and not allow transmission of diffracted X-rays. You can calculate absorption for capillary based samples here:

https://11bm.xray.aps.anl.gov/absorb/absorb.php

Patterns with high absorption have a rising background at low and high 2theta values, giving the pattern a characteristic “smile”, relative intensities will be affected and in extreme cases, there may be shifts in 2theta. You can dilute your samples with either a known phase or an amorphous substance to help address this. Ideally, to ensure that absorption corrections can be applied to relative intensities, the absorption coefficient muR, should be <1.

As a general rule, higher energy X-rays are more penetrating so a higher energy can help absorption effects.

Example of a pattern with absorption effects.

Example of a pattern with absorption effects.

 

Avoiding Fluorescence/ Absorption edges

The Mythen detector used at PD is very sensitive to fluorescence and so data must be collected far enough away from an edge to avoid the detector being swamped by fluorescence.

The data should be collected either below an edge, or at least 6keV above an edge.

Examples:

  • For samples containing Fe, where K-alpha is ~7.1keV, samples must be collected at energies >13keV

  • For samples containing Zn, where K-alpha is at ~9.7keV, samples must be collected at energies>16keV

  • For samples containing Zr, where k-alpha is at ~17.99keV, samples should be collected at <18keV

The position of edges can be calculated using this tool:

https://physics.nist.gov/PhysRefData/FFast/html/form.html

by plotting the Mass Absorption Coefficient between 10 and 22 keV to cover the beamline energy range.

An example of a Y containing compound collected at 18keV

An example of a Y containing compound collected at 18keV

Compared to the same samples at 15keV:

 

Beamline flux

The flux of the beamline is maximised between 12 - 16 keV, so you might need to count longer at higher energies as can be seen from this figure.

 

Large unit cells with low angle peaks

In this case, you may want to balance your energy choice to ensure that any expected low angle peaks are at a reasonable 2theta value >~3 or 4° to ensure they are not hidden by the beamstop.