Bending magnet

Available energy range

8 - 22 keV (0.5956 - 1.6 54 angstroms)

 Mirrors

Vertical collimating mirror (VCM)

  • collimates beam

2 Modes:

  • Flat for experiments in Hutch C

  • Toroid for focused beam in Hutch B

 

Vertically focusing mirror (VFM)

  • Focuses beam vertically

2 Modes:

  • Flat for experiments in Hutch B

  • Toroid for focused beam in Hutch C

Double crystal monochromator (DCM)

Si(111) flat crystal pair

Nominal beam size at sample

Mode 1

VCM flat & VFM Flat, 3 mm (H) x 1 mm (V)

Mode 2

Hutch B - VCM toroid & VFM flat, 1mm (H) x 1 mm (V)

Mode 3

Hutch C - VCM flat & VFM toroid, 1mm (H) x 1 mm (V)

🔷 For measurements with Diamond Anvil Cells, this will be reduced further as required.

Detectors

Mythen-II

High resolution position sensitive detector - 80° 2Theta

Mar345 Image Plate

2D detector (angular range varies depending on detector distance)

Varex XRPad 4343F

2D detector (angular range varies depending on detector distance)

...